71
Specifications
12.9 Temperature (Dual Channel)
12.10
CO
2
Method
Thermistor (YSI 400 series compatible)
Accuracy
±0.1°C ± 1 digit
Range
20°C - 45.5°C
Display
T1 or T1+T2+
Trends
1, 8 or 24 hour trend of T1 and T2
Alarms
High and Low temperature on T1 and T2.
Core technology
Respironics LoFlo™
Mode of sampling
Sidestream
Principle of
operation
NDIR single beam dual wavelength optics
Initialisation time
Capnogram displayed in less than 20s
(ambient temp of 25ºC), full specification within
2 minutes
CO
2
measurement
range
0 to 150mmHg
0 to 19.7%
0 to 20kPa
CO
2
Calculation
method
BTPS (Body Temperature Pressure Saturated)
CO
2
Response
time
<3s - Includes transport and rise time
CO
2
Resolution
0.1mmHg - 0 to 69mmHg
0.25mmHg - 70 to 150mmHg
CO
2
Accuracy
0 - /- 2mmHg
41 - /- 5% of reading
71 - 1/- 8% of reading
101 - 1/- 10% of reading
Above 80 breaths per /- 12% of
reading
CO
2
Stability
<0.8mmHg in 4 hrs
Respiration rate
2 to 150 breaths per minute
Summary of Contents for Smartsigns Compact SC1000