28
ECG Monitoring
Do not allow the ECG cable to place stress on to the ECG
electrodes, as this will produce poor quality recordings.
NOTE - If reusable or non-gelled electrodes are used, apply gel to
the centre of the electrodes.
If pre-gelled disposable electrodes are used, peel the backing from
the electrode adhesive and apply the electrodes according to the
manufacturers instructions.
The patient cable must be inserted into the green ECG socket on
the system.
12.4 Setting up the ECG waveform
ECG SETUP
Auto Limit
150
50
+
-
+
-
HR
High
Low
ON
II
Parameter
Lead
Gain
Mode
Filter
QRS Beep
DUAL
OFF
OFF
50Hz, 60Hz
ON
ON
OFF
MONITOR
x1
1MV
x2, x4, x
1
2
DIAGNOSTIC
OFF
Press anywhere on the ECG display area to display the ECG set-up
options: -
This switches the ECG monitoring function ON or OFF.
Use the touch screen to the adjust setting. If the ECG channel is
switched OFF, OFF is displayed in large letters on the right hand
side of the screen.
Press to exit from this function.
Parameter