58
CO
2
Monitoring
18.9 CO
2
Error Messages
ZERO the system if prompted by the
monitor.
Apply the sampling line to the proximal
end of the airway circuit as shown.
After a short period, the capnogram
will be displayed on the monitor's
screen.
18.8 CO
2
Alarm Settings
Press the CO
2
area on the screen to display the CO
2
set up options.
Use the + or - softkeys to adjust the setting as necessary.
Furthermore, the user can set AUTO LIMIT which automatically
adjusts the alarms to match patient status.
Close the set up screen to save settings.
Connecting to
integrated airway
adaptor
Connecting to integrated
airway adaptor via elbow
type of connector
Error Message
Rectification
Sensor Over Temp
Ensure sensor is not exposed to extreme heat (eg.
heat lamp).
Sensor Fault
Check sensor is properly plugged in. Re-insert if
necessary.
Zero Required
Check airway adaptor and clean if necessary.
Out of Range
CO
2
out of range
Check Airway adaptor Check and clean airway adaptor if necessary.
Check Sampling Line
Check that sampling line is not occluded or kinked.
System Error
Contact Huntleigh Healthcare Service Dept.
Error XXX
Contact Huntleigh Healthcare Service Dept.
Summary of Contents for Smartsigns Compact SC1000
Page 1: ...SC1000 Smartsigns Compact...