51
Invasive Blood P
ressure Monitoring
17.5 IBP set up options
Touch the appropriate IBP area of the screen to display the set up
screen: -
IBP2 SETUP
ON
ZERO
Parameter
Layout
- - - / - - -
(- - -)
- - -
Auto Limit
OFF
- - -
- - -
20.0
5.0
+
-
+
-
High
Low
20.0
5.0
+
-
+
-
High
Low
20.0
5.0
+
-
+
-
High
Low
SYS
ON
DIA
ON
MEAN
ON
OFF
OFF
OFF
IBP1 SETUP
ON
ZERO
Parameter
Layout
- - - / - - -
(- - -)
- - -
Auto Limit
OFF
- - -
- - -
160
50
+
-
+
-
High
Low
160
50
+
-
+
-
High
Low
160
50
+
-
+
-
High
Low
SYS
ON
DIA
ON
MEAN
ON
OFF
OFF
OFF
This switches the IBP 1 monitoring function ON or OFF.
If the IBP 1 channel is switched OFF, 'OFF' is displayed in large
letters on the right hand side of the screen.
Press to exit from this function.
Parameter
Press the
TAB to zero the transducer.
A two second delay occurs while the zero adjustment takes place.
If an error message appears, check that the pressure transducer is
not pressurised and that the electrical connectors are correctly
fitted.
Zero
ZERO