5.14 Buzzer Replacement
Disconnect connector CON1. Cut wires from existing buzzer mounted on base.
Fit replacement buzzer onto a clear base area close to the front of the pump
ensuring that the lead is able to reach and fit onto
CON1
.
Use Cyanoacrylate 495 adhesive to bond the sounder’s copper side to the plastic
case.
Ensure the sounder is firmly attached by applying pressure for 10 seconds.
Note: Sounder clarity and volume are dependant on the quality of this bonding
so ensure sounder output is checked by forcing the pump into a fault condition
in order to check sounder output.
5.15 Hang Strap Assembly
Place the unit upside down on a flat, clean surface. Unscrew and remove two
retaining screws from the hang strap plates anchoring the assembly to the base
of the pump unit. Lift the assembly away from the unit and replace it.
247901
5-7
Issue 1
March 1998
Summary of Contents for Flowtron AC550
Page 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Page 2: ...247901 Issue 1 Mar 1998...
Page 4: ...247901 ii Issue 1 March 1998...
Page 6: ...247901 iv Issue 1 March 1998...
Page 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Page 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Page 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Page 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Page 16: ...247901 1 8 Issue 1 March 1998...
Page 18: ...247901 2 2 Issue 1 March 1998...
Page 30: ...247901 4 8 Issue 1 March 1998...
Page 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Page 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Page 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Page 38: ...247901 5 8 Issue 1 March 1998...
Page 40: ...247901 6 2 Issue 1 March 1998...