4.1.2
General Service - Pump
Initial Assessment (Visual)
Check:
a)
Enclosure condition
b)
Fittings for damage and security
c)
Labelling is secure and legible
Open pump and check:
d)
Wiring security and terminations are secure
e)
Tubing security and integrity
f)
Component fastenings are secure
g)
General condition inside is clean
Open compressor and check:
h)
Bellows assembly condition
i)
Silencer bag weld seams
General:
j)
Check power cord assembly for security
k)
Clean condition of pump
If any items in a) - j) appear defective, replace using an appropriate spare part
and record action on the service record sheet.
Re-assemble pump unit completely, run for a minimum of 1 hour and check
calibration. Adjust if required.
Note:
It is important to ensure all tools used for servicing are kept in good
condition, and test rigs for measuring flow and pressure are regularly
calibrated to a reference master gauge calibrated to a National Standard.
4.2
Pressure Output and Calibration Checks
A calibrated Manometer/Pressure Meter with better than 0.5 mmHg accuracy.
In normal operation, when the display reads 40, the knob points should be
somewhere within the marked 40 mmHg arc. If not, the pump should be
returned for re-calibration. The error could be due to :
a)
Inaccuracy of the pressure setting control - see Section 4.3
Pressure Control Calibration.
247901
4-2
Issue 1
March 1998
Summary of Contents for Flowtron AC550
Page 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Page 2: ...247901 Issue 1 Mar 1998...
Page 4: ...247901 ii Issue 1 March 1998...
Page 6: ...247901 iv Issue 1 March 1998...
Page 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Page 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Page 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Page 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Page 16: ...247901 1 8 Issue 1 March 1998...
Page 18: ...247901 2 2 Issue 1 March 1998...
Page 30: ...247901 4 8 Issue 1 March 1998...
Page 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Page 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Page 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Page 38: ...247901 5 8 Issue 1 March 1998...
Page 40: ...247901 6 2 Issue 1 March 1998...