3
TROUBLESHOOTING
The
Flowtron
®
Excel
pump unit has both an audible and a visual alarm. If a
problem occurs, the system will sense the fault and briefly flash a message on
the display. It will not alarm immediately in order to screen out false alarms
due to normal patient movement. If the same fault continues on ten successive
inflations, the audible alarm will sound and a flashing message will remain on
the display until corrective action is detected.
Note: The exception is the “F” system alarm which will be activated on sensing
the fault on the first occasion.
3.1
Preliminary Checks
HI
or
Lo
faults have probably been caused by garment or tubing problems.
Garment faults should be checked as below.
Display
Problem
Corrective Action
1.
Loose fitting garment.
2.
Hose disconnected at
garment.
3.
Garment leak.
4.
Low pressure.
1.
Rewrap snugly.
2.
Check the hose connection
at garment end.
3.
Check garment and
replace if faulty.
4.
Refer to service.
1.
Hose kinked causing a
blocked tube.
2.
Hose disconnected at pump.
3.
Single garment attached
without pressing “Single Leg”
button.
1.
Check hoses for kinks
or obstructions.
2.
Check the hose connection at
pump outlet.
3.
Press “Single Leg” button if
only one garment is used.
Pump failure.
DO NOT USE PUMP
.
Refer to service.
In the unlikely event of a monitoring system electronics failure, the red LED on
the top panel will flash rapidly to indicate monitoring system failure.
If garment faults have been eliminated and the fault persists, proceed to Service
Mode Checks.
247901
3-1
Issue 1
March 1998
Summary of Contents for Flowtron AC550
Page 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Page 2: ...247901 Issue 1 Mar 1998...
Page 4: ...247901 ii Issue 1 March 1998...
Page 6: ...247901 iv Issue 1 March 1998...
Page 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Page 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Page 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Page 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Page 16: ...247901 1 8 Issue 1 March 1998...
Page 18: ...247901 2 2 Issue 1 March 1998...
Page 30: ...247901 4 8 Issue 1 March 1998...
Page 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Page 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Page 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Page 38: ...247901 5 8 Issue 1 March 1998...
Page 40: ...247901 6 2 Issue 1 March 1998...