3.2
Service Mode Checks
Service mode gives additional information on the type of alarm fault.
To access the service mode, hold the Single Leg
button down for
approximately 13 seconds. The unit will then ‘beep’ and the red LED on the top
panel will be constantly displayed.
The main display will now indicate the actual pressure to 1 mmHg resolution
and also display the detailed fault codes.
Fault codes and counter measures are described in the following table (see
pages 3-3 and 3-4).
3.3
Loss of Electrical Power
Power loss can be due to the following :
1. No electricity supply at wall outlet.
2. Faulty wall outlet.
3. Damaged cord or cord plug.
4. Defective fuse link.
5. Damaged internal wiring.
6. Cord plug inadvertently removed from wall outlet.
3.4
247901
3-2
Issue 1
March 1998
Summary of Contents for Flowtron AC550
Page 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Page 2: ...247901 Issue 1 Mar 1998...
Page 4: ...247901 ii Issue 1 March 1998...
Page 6: ...247901 iv Issue 1 March 1998...
Page 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Page 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Page 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Page 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Page 16: ...247901 1 8 Issue 1 March 1998...
Page 18: ...247901 2 2 Issue 1 March 1998...
Page 30: ...247901 4 8 Issue 1 March 1998...
Page 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Page 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Page 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Page 38: ...247901 5 8 Issue 1 March 1998...
Page 40: ...247901 6 2 Issue 1 March 1998...