3.0.0.1
Fault Codes and Counter Measures
247901
3-3
Issue 1
March 1998
MESSAGE User
Mode
Service Mode
Fault
Detection
Protocol
Possible Causes
User
Faults
Possible Causes
Pump
Faults
Effect
Reset Method
Lo
L4
A
single
event
where
no
pressure
greater
than
4mmHg,
is
detected
for
135
secs.,
which
then
activates
the
alarm.
During
this
period,
a
Lo
message
briefly
appears
on
the
display,
at
30
sec.
intervals,
but
not
necessarily
in
synchronisation
with
the
air
output.
Both
supply
tubes
disconnected
at
the
garment
connectors.
Rotary
valve
failed
in
the
closed
position
Compressor
failure.
Pressure
control
valve
fault.
No
inflation
of
both
garments.
1
normal
cycle,
or
power
reset.
FF
3
Pressure
greater
than
72mmHg
detected
for
6
secs.,
which
then
activates
the
alarm.
N/A
Pressure
control
can
be
set
too
high,
due
to
incorrect
adjustment. Monitoring
PCB
system,
pressure
display/calibration fault.
Garment pressure
is
higher
than
the
therapeutic recommendation s
Power
reset
only.
FF
4
Pressure
greater
than
15mmHg
detected
for
40
secs.,
which
then
activates
the
alarm.
Note
:
Constant
high
priority
alarm
note.
N/A
Valve
failed
in
the
fully
open
position.
Constant inflation
of
the
garment,
giving
a
tourniquet
effect.
Power
reset
only.
Summary of Contents for Flowtron AC550
Page 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Page 2: ...247901 Issue 1 Mar 1998...
Page 4: ...247901 ii Issue 1 March 1998...
Page 6: ...247901 iv Issue 1 March 1998...
Page 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Page 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Page 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Page 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Page 16: ...247901 1 8 Issue 1 March 1998...
Page 18: ...247901 2 2 Issue 1 March 1998...
Page 30: ...247901 4 8 Issue 1 March 1998...
Page 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Page 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Page 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Page 38: ...247901 5 8 Issue 1 March 1998...
Page 40: ...247901 6 2 Issue 1 March 1998...