29
Oper
ation
Foot Position
Adult
18cm - 28cm (7.1
Ý - 1
1.0
Ý
)
Min
Ensure patients heel
is resting on the
couch. Do not rest
the leg on the cuff
as this may affect
measurement result.
Incorrect Cuff Placement
Huntleigh Healthcar
e Ltd
Diagnostic P
roducts Division
35 Portmanmoor Road
Cardiff, CF24 5HN. UK
Tel: +44(0) 29 2048 5885
Email: sales@huntleigh-
diagnostics
.co.uk
www.huntleigh-
diagnostics
.com
ACC-VAS-023
Made in UK
0088
LATEX
PVC
Ensure the cuffs are not
rotated around the limb.
Measurement error or
failure to take correct
measurement may
occur if cuffs are fi tted
incorrectly
Adult
18cm - 28cm (7.1
Ý - 1
1.0
Ý)
Max
Min
Ensure the cuffs are fi tted
in the correct orientation.
Measurement error or
failure to take correct
measurement may
occur if cuffs are fi tted
incorrectly
Note:
Take great care to ensure that the tubing is not kinked or
obstructed in any way. Do not touch cuffs or tubing when
measurement is in progress.
Note:
The patient must lie supine, be relaxed, remain still, and refrain
from talking, coughing etc.
Note:
Always brief the patient before the test, explaining that the cuffs
will tighten, and that the test will take approximately 3 minutes
to
complete.
Summary of Contents for dopplex ABIlity
Page 1: ......
Page 56: ...771528 4 ENGLISH USA 2011...