34
Tr
oubleshooting
13.1 Error Messages
Cuff Not Connected
If the user fails to connect one or more of the
cuffs to the unit or the cuffs are not tensioned
correctly, one of the following messages is
displayed:-
•
Left / Right Arm Cuff - Not Connected
•
Left / Right Ankle Cuff - Not Connected
•
Arm Cuffs / Ankle Cuffs - Not Connected
Left Arm Cuff - Not Connected
Refer To User Manual
Recommended Action:
Check that the cuffs are properly connected, and are
fi tted snugly. Refer to section 12.4.3.
Cuff Air Leak
If an air leak is detected in the cuffs or tubing,
one of the following messages is displayed:-
•
Left / Right Arm Cuff - Leak Detected
•
Left / Right Ankle Cuff - Leak Detected
•
Arm Cuffs / Ankle Cuffs - Leak Detected
Left Arm Cuff - Leak Detected
Refer To User Manual
Recommended Action:
Run the Cuff Leak test in Section 13.2.
Infl ation Problem
If any cuff fails to infl ate correctly, one of the
following messages is displayed:-
•
Left / Right Arm Cuff - Infl ation Problem
•
Left / Right Ankle Cuff - Infl ation Problem
•
Arm Cuffs / Ankle Cuffs - Infl ation Problem
Left Arm Cuff - Inflation Problem
Refer To User Manual
Recommended Action:
Check cuffs are fi tted correctly and tubing is
not kinked or obstructed. If problem persists, refer to Service Manual for
diagnostic test procedures.
Excess Pressure
If excess pressure is detected in any cuff, one of
the following messages is displayed:-
•
Left / Right Arm Cuff - Excess Pressure
•
Left / Right Ankle Cuff - Excess Pressure
•
Arm Cuffs / Ankle Cuffs - Excess Pressure
Left Arm Cuff - Excess Pressure
Refer To User Manual
Recommended Action:
Ensure the cuffs are not obstructed and are free
to infl ate properly. Do not squeeze the cuffs during tests, or allow them to
become trapped, e.g. against a wall or similar hard surface.