ColorFlex EZ User’s Manual Version 2.
2
52
Press
DIAGS
MENU
to return to the Diagnostics menu. Press
MAIN MENU
to return
to the main menu.
•
Signal Levels
The Signal Levels diagnostic prompts to place a sample at the measurement port.
Normally this will be the White Tile, but other instructions may be provided by
HunterLab’s
Service Support staff. Place the White Tile at the port and press GO.
The sample and monitor channels are plotted on a graph for examination by
HunterLab’s
Service Support staff.
Figure 76. Signal Levels
Press
READ
to read again. Press
DIAGS MENU
to return to the Diagnostics menu.
Press
MAIN MENU
to return to the main menu.
•
Self-Test
The
DIAGNOSTICS MENU
contains a fourth function named
SELF TEST
. When this
function is selected, communications tests between the ColorFlex
EZ’s subsystems
are run. If successful, the versions of the firmware present would be displayed. Also
shown is the value of the checksum of the instrument’s calibration data as four
hexadecimal digits.
Figure 77. Self-Test
Summary of Contents for ColorFlex EZ
Page 6: ...ColorFlex EZ User s Manual Version 2 2 6...
Page 16: ...ColorFlex EZ User s Manual Version 2 2 16...
Page 30: ...ColorFlex EZ User s Manual Version 2 2 30...
Page 36: ...ColorFlex EZ User s Manual Version 2 2 36 Figure 54 Date Time Screen...
Page 40: ...ColorFlex EZ User s Manual Version 2 2 40...
Page 46: ...ColorFlex EZ User s Manual Version 2 2 46...
Page 54: ...ColorFlex EZ User s Manual Version 2 2 54...
Page 58: ...ColorFlex EZ User s Manual Version 2 2 58 Regulatory Notice...