27
Standard Values
The Standard Values parameter is only relevant when the Standard Type is physical, numeric, or
hitch. When Standard Values is highlighted, press the
RIGHT
arrow button to move to the
STANDARD VALUES
configuration screen.
Figure 38. Standard Values
•
For a
PHYSICAL
standard, place the standard at the sample port and press the
GO
button (Read) to read the standard and place its measurement values (in the
selected color scale) in the product setup. Then press the
LEFT
arrow button (Done)
to return to the product setup configuration screen.
•
For a
NUMERIC
standard, change the highlighted digit of the selected color scale
value using the
UP
and
DOWN
arrow buttons. When the desired digit is shown,
press
NEXT
to move to the next digit. When all digits are set as desired, press
DONE
to return to the product setup configuration screen.
Figure 39. Enter a Numeric Standard
•
For a
HITCH
standard, place the hitch standard (with known values from a reference
instrument) at the sample port and press the
GO
button (Read) to read the standard
and place its measurement values (in the selected color scale) in the product setup.
Figure 40. Hitch Standard
Summary of Contents for ColorFlex EZ
Page 6: ...ColorFlex EZ User s Manual Version 2 2 6...
Page 16: ...ColorFlex EZ User s Manual Version 2 2 16...
Page 30: ...ColorFlex EZ User s Manual Version 2 2 30...
Page 36: ...ColorFlex EZ User s Manual Version 2 2 36 Figure 54 Date Time Screen...
Page 40: ...ColorFlex EZ User s Manual Version 2 2 40...
Page 46: ...ColorFlex EZ User s Manual Version 2 2 46...
Page 54: ...ColorFlex EZ User s Manual Version 2 2 54...
Page 58: ...ColorFlex EZ User s Manual Version 2 2 58 Regulatory Notice...