![HunterLab ColorFlex EZ User Manual Download Page 51](http://html1.mh-extra.com/html/hunterlab/colorflex-ez/colorflex-ez_user-manual_2186985051.webp)
ColorFlex EZ User’s Manual Version 2.
2
51
Figure 73. White Repeatability Test Results
Press
to print the result (if a printer is connected to the instrument’s USB
port). An example printout is shown.
Figure 74. Printout of Repeatability Test
Press
DIAGS
MENU
to return to the Diagnostics menu to run further tests. Press
MAIN MENU
to return to the main menu.
•
Green Tile Check
The first Green Tile Check screen allows you to compare the values in the ColorFlex
EZ’s memory to the values read at factory on
the back of the diagnostic Green Tile. If
the values match, press the
GO
button to continue. If the values in the instrument’s
memory require revision, change the highlighted digit of the selected color scale
value using the
UP
and
DOWN
arrow buttons. When the desired digit is shown,
press
NEXT
to move to the next digit. When all digits are as desired, press
GO
to
continue. Press
BACK
to cancel the test.
Figure 75. Green Tile Values
Standardize the instrument as prompted and when complete, place the diagnostic
Green Tile at the sample port. Press the
GO
button. One reading of the Green Tile is
made, and the result with a pass/fail rating is shown. To pass, dX, dY, and dZ must all
be less than or equal to 0.30
Press
to print the result (if a printer is connected to the instrument’s USB
port).
Summary of Contents for ColorFlex EZ
Page 6: ...ColorFlex EZ User s Manual Version 2 2 6...
Page 16: ...ColorFlex EZ User s Manual Version 2 2 16...
Page 30: ...ColorFlex EZ User s Manual Version 2 2 30...
Page 36: ...ColorFlex EZ User s Manual Version 2 2 36 Figure 54 Date Time Screen...
Page 40: ...ColorFlex EZ User s Manual Version 2 2 40...
Page 46: ...ColorFlex EZ User s Manual Version 2 2 46...
Page 54: ...ColorFlex EZ User s Manual Version 2 2 54...
Page 58: ...ColorFlex EZ User s Manual Version 2 2 58 Regulatory Notice...