ColorFlex EZ User’s Manual Version 2.
2
53
The value of the checksum of the calibration data should remain constant unless the
White Tile
’s ass
igned reflectance is updated or the instrument is serviced. A record
of the checksum’s value
should be kept. If it changes for unexplained reasons then
the instrument may not function correctly.
Figure 78. Self-Test Results
Press
DIAGS MENU
to return to the Diagnostics menu. Press
MAIN MENU
to return
to the main menu.