P/N: 7000717 Rev A, CO-148, Released Date: 2021-01-28;
©2021, IRRAS USA, Inc., All rights reserved
Page
17
of
38
Run Mode
Set appropriate intracranial pressure alarm limits on the bedside patient monitor.
Note:
For continuous monitoring of intracranial pressure intra-hospital transport, check the battery
status lights to confirm that the battery is fully charged before unplugging.
If, in Run-Mode, the catheter is disconnected and then reconnected, the Hummingbird ICP
Control Module will sense this event and require the user to re-prime the system.
The Hummingbird ICP Control Module has BLUE indicator lights for the priming operation. A
pulsating light indicates the Hummingbird ICP Control Module is being primed.
Solid
BLUE
lights are displayed when the Hummingbird ICP Control Module is in Run Mode.
The PRIME button can be pressed when in Run Mode to check the system and to ensure the
device is working accurately.
Note:
After the initial prime sequence, the Hummingbird ICP Control Module will automatically
perform a priming sequence three times within the first two hours: at 5 minutes, at 30
minutes and at 90 minutes and 60 minutes thereafter. This is designed to check the
integrity of the Hummingbird ICP Control Module system. During the priming sequence,
the ICP waveform on the patient monitor will be a flat line reading approximately 0 mmHg
until priming is completed (note: this shall take less than 35 seconds).
After this initial check, the Hummingbird ICP Control Module will typically re-prime
automatically every hour.
Patient Transportation
Disconnecting the Patient Cable Module from the ICP Catheter Connector to transport
the patient
1.
Leave the ICP Control Module on Utility Pole and Monitor Cable attached to the IBP
port on the patient monitor.
2.
Disconnect the ICP Catheter Connector from the Patient Cable Module.
3.
Place the ICP Connector Cap onto ICP Catheter Connector
4.
Insert the HICP200 Plug into the Patient Cable Module
5.
Place the Patient Cable Module into cable hook at base of ICP Control Module.
The Hummingbird ICP Control Module with Monitor Cable and
Patient Cable Module is contraindicated for use in a Magnetic
Resonance (MR) environment.