Caution
Only use the cleaning agents listed in this section for cleaning and disinfecting the
Hummingbird ICP Control Module and its components. Using solvents or cleaning agents not
listed in the cleaning guidelines may damage the plastic exterior of the Hummingbird ICP
Control Module.
Warning
Do not autoclave, use automated cleaning methods, or immerse the ICP Control
Module in liquid as damage may occur. If the monitor is exposed to liquids, remove
the AC Power Supply, dry the unit thoroughly, and send to biomed staff for evaluation
before reapplying power.
Item
Guideline
Preparation for
Cleaning
Ensure that all the connections between the cables, catheter, AC Power
Supply, and monitor have been removed before cleaning the
Hummingbird ICP Control Module.
Recommended
Manual
Cleaning
Method
Using either 70% IPA with a lint free wipe or a Diversey Oxivir-Tb
Wipes®, thoroughly wipe all surfaces at least three (3) times and then
inspect the surfaces for visible residues. If residues remain, use a new
Diversey Oxivir-Tb Wipes® or lint free wipe soaked with 70% IPA and
continue wiping the surfaces until they are visibly free of residues.
Recommended
Disinfection
Agents
Hummingbird ICP Control Module has been tested and shown to
withstand exposure to 70% IPA and Diversey Oxivir-Tb Wipes®. If used
follow standard hospital procedures and
manufacturer’s instructions for
use.
Inspection
After each reprocessing event, visually inspect the ICP Control Module
and any of its system components for any wear and tear.
Containment
and
Transportation
It is recommended that the ICP Control Module and any of its system
components be cleaned as soon as is reasonably practical after use.
Cleaning Patient Cable Module
1. Clean the outside of Patient Cable Module with 70% Isopropyl Alcohol (IPA) with a
lint free wipe or a Diversey Oxivir-Tb Wipes®.
2. Moisten a swab with 70% IPA.
3. Insert the swab inside the Patient Cable Module connector.
4. Rotate the swab inside the connector.
5. Insert an HICP200 Plug into the Patient Cable Module
Note About Hummingbird Neuromonitoring Catheters
Hummingbird Neuromonitoring catheters are intended for single use only and cannot be
resterilized. Hummingbird Neuromonitoring catheters have not been designed for
multiple use and have not been tested for resterilizability in the hospital setting. Refer to
the Hummingbird H110 and H610 IFU
’s for additional information.