During immunity testing, the Hummingbird ICP Control Module maintained its ability to
monitor pressure within specification or issued an alarm to advise the user that the data
was not reliable. Faults could be corrected by cycling the power.
Warning
Use of accessories and cables other than those specified by IRRAS USA, Inc. may result
in increased emissions or decreased immunity of the Hummingbird ICP Control Module.
Only use IRRAS USA, Inc. authorized accessories and cables.
Guidance and manufacturer’s declaration – electromagnetic immunity
The
Hummingbird ICP Control Module
is intended for use in the electromagnetic environment
specified below. The customer or the user of the
Hummingbird ICP Control Module
should
assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2 kV, ±4 kV, ±8
kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
±2 kV, ±4 kV, ±8
kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The
Hummingbird ICP Control Module
is intended for use in the electromagnetic environment
specified below. The customer or the user of the
Hummingbird ICP Control Module
should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11 (150
kHz
– 30 MHz)
Class A
The
Hummingbird ICP Control Module
uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11 (30
MHz
– 1000
MHz)
Class A
Harmonic
emissions
IEC 61000-3-2
Complies
The
Hummingbird ICP Control Module
is suitable for use
in all establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies