To reduce the risk of electric shock, do not disassemble the Hummingbird ICP Control
Module. Refer all servicing to qualified service personnel at IRRAS USA, Inc.
To prevent damage to the precision equipment or personnel, only use the AC
Power Supply provided by IRRAS USA, Inc. The AC Power Supply is identified
and labeled to ensure proper identification. Reference the section Abbreviations
and Symbols to correctly identify the AC Power Supply.
Danger - Possible explosion hazard if used in the presence of flammable anesthetics.
During initial setup, 10 hours charging on AC power is required to obtain full capacity.
When relying on battery power, ensure adequate battery capacity for the intended
duration.
Routinely check the performance and desired capacity of the battery prior to and during
use.
Connect the monitor to an AC Power Supply immediately if the low battery alarm is
activated.
To ensure safe operation:
Use of accessories and cables other than those specified by IRRAS USA, Inc.
may result in increased emissions or decreased immunity of the Hummingbird
ICP Control Module. Only use IRRAS USA, Inc. authorized accessories and
cables.
Only use the Hummingbird ICP Control Module with the cables, catheter
systems, and authorized accessories described in these instructions for use.
Use of accessories and cables other than those specified by IRRAS USA, Inc.
may result in increased emissions or decreased immunity of the Hummingbird
ICP Control Module.
Never modify the Hummingbird ICP Control Module or its cables.
Only use IRRAS supplied accessories on the Hummingbird ICP Control Module. This
applies in particular to catheters, catheter cables, battery, and AC Power Supply.
Do not autoclave, use automated cleaning methods, or immerse the Hummingbird ICP
Control Module, Monitor Cable or Patient Cable Module in liquid as damage may occur. If
the Hummingbird ICP Control Module, Monitor Cable or Patient Cable Module is exposed
to liquids, turn off the unit, remove the AC Power Supply, dry the unit thoroughly, and
send to biomed staff for evaluation before reapplying power.