24
www.humancaregroup.com
Guidance and manufacturer’s declaration
ENG
Guidance and manufacturer’s declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The
customer or the user of the Human Care lift should assure that it is used in such an environment.
Emission test
Compliance
Electro -
magnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Human Care lift uses RF energy only for
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The Human Care lift is suitable for use
in all establishments including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The
customer or the user of the Human Care lift should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electro -
magnetic environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic
tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Electrical
fast
transient /
Burst
IEC 61000-
4-4
+/- 2 kV for
power supply
lines
+/- 1 kV for
input/output
lines
+/- 2 kV for
power supply
lines
n/a. for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment.