47
Three-phase energy meter field calibrator
1. Open the instrument, and press
【
SWITCH TO
】
to enter the "double meter" check interface;
2. Press
【
SET
】
meter 1 Constant, and enter "08000"; press
【
↓
】
to lock, and move the cursor
to N (number of turns) setting item;
3. Input "02" to N (turns) setting item; press
【
↓
】
to lock and move the cursor to the active and
reactive power setting term;
4. Press
【
← →
】
on the active and reactive power setting item and select P; press
【
↓
】
, and
move the cursor to the photoelectric and manual setting item;
5. Select photoelectric on the photoelectric and manual setting item through
【
← →
】
; press the
constant of
【
↓
】
meter 2;
6. On THE constant setting in meter 2, input "08000"; press
【
↓
】
and move the cursor to N
(number of turns) setting item;
7. On the N (number of turns) setting item, input "03"; press
【
↓
】
and move the cursor to the
active and reactive power setting term;
8. On the active and reactive power setting item, through
【
← →
】
select "Q"; press
【
↓
】
and
move the cursor to the photoelectric and manual setting item;
9. On the optoelectronic and manual setting item through
【
← →
】
select the "photoelectric";
press
【
↓
】
and move the cursor to the input setting item;
10. On the input setting item, select "5A Clip-on ammeter" through
【
← →
】
key;
11. Press
【
】
key to start calibration.
Summary of Contents for HZGC3521 Series
Page 8: ...7 Three phase energy meter field calibrator 1 2 Introduction to top panel structure...
Page 9: ...8 Three phase energy meter field calibrator 1 3 Introduction to right panel structure...
Page 10: ...9 Three phase energy meter field calibrator 1 4 Back Introduction...
Page 31: ...30 Three phase energy meter field calibrator 2 7 Description of Function Keys...
Page 40: ...39 Three phase energy meter field calibrator...
Page 70: ...69 Three phase energy meter field calibrator...
Page 79: ...78 Three phase energy meter field calibrator...