ECLIPSE - HT9025
EN - 31
5.11. INRUSH CURRENT MEASUREMENT
CAUTION
Make sure that all the test leads are disconnected from the meter terminals
for current measurement.
Fig. 33: Use of the instrument for AC Inrush current measurement
1. Select positions
60A
,
600A
or
1000A
2. Press the
MODE/VFD
key to select
“
A
”
measurement
3. Long press the key
/INRUSH
to activate the “INRUSH” function. The “
” and “- - - -“
symbols appears at display (see Fig. 33 – left side)
4. Put the conductor to be tested inside to the center of clamp jaw (see Fig. 33 – right
side) in order to perform accurated measurements. Consider the notches on jaws as
reference (see Fig. 3)
5. The display shows the value of inrush current.
6. If the display shows the message
“
OL
”
, the maximum measurable value has been
reached.
7. To use the HOLD function, see § 4.2
8. To use the internal functions, see § 4.3
9. To save the measured result, see § 4.3.4