DIAGnostic:OUTPut:ALL[:STATe]
DIAGnostic:OUTPut:ALL[:STATe] <
state
>
opens or closes all channel output
relays at once.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
<
state
>
boolean
OFF | ON | 0 | 1
0
Comments
•
This command performs the same function as the
OUTPut
n
[:STATe]
command,
but enables or disables all outputs simultaneously.
•
Related Commands:
OUTPut
n
[:STATe]
DIAGnostic:SOURce
n
:FUNCtion:MODE
DIAGnostic:SOURce
n
:FUNCtion:MODE <
mode
>
changes the output mode of the
channel without disabling the channel output.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n
keyword
1 to 8 or 1 to 16
1
<
mode
>
discrete
CURRent or VOLTage
none
Comments
•
In the keyword,
SOURce
n, the valid range of n is:
1 to 8 for
E1418A, E1418A Option 002
1 to 16 for
E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
•
Use this command with caution. The output voltage or current is not
predictable during the mode change and transients will occur.
•
This command performs the same function as the
SOURce
n
:FUNCtion:MODE
command, but does not disable the output during the mode change. This
command can reduce the settling time required to disable and enable the outputs.
•
Following a mode change with this command, the appropriate calibration constants
are used, the channel mode is changed, and the output is set to either 0 V or 0 A.
•
This command will generate an error if the trigger system is waiting for a trigger
event when this command is received.
•
This command is affected by the state of the P/J jumper on the terminal module.
If the P/J jumper is not set to Program Selectable, this command will generate an
error. This jumper is described on page 31.
DIAGnostic
77 HP E1418A SCPI Command Reference
Chapter 3
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...