Channel Trigger
Register
The Channel Trigger Register reports the trigger status for each channel.
Address
Base + 18
16
Bit
15
14
13
12
11
10
09
08
07
06
05
04
03
02
01
00
READ
Ch
16
TM
Ch
15
TM
Ch
14
TM
Ch
13
TM
Ch
12
TM
Ch
11
TM
Ch
10
TM
Ch
09
TM
Ch
08
TM
Ch
07
TM
Ch
06
TM
Ch
05
TM
Ch
04
TM
Ch
03
TM
Ch
02
TM
Ch
01
TM
Ch XX TM: A one (1) in a field indicates a value has been written to the
channel Main DAC Triggered Register, but the output is not enabled until the
trigger is received. A zero (0) in a field indicates a trigger will have no effect
on the channel. The trigger events are set in the Trigger Control Register.
Notes
You can output an immediate value on a channel that has been set for
triggered operation by writing a value to the Main_DAC Immediate
Register. The trigger mode for that channel is changed to the immediate
write-thru mode.
You can set a value on a channel to be applied at the next trigger event by
writing a value to the Main_DAC Triggered Register. The trigger status for
that channel will be changed to triggered.
Channels 09 through 16 will return ones (1) if the module is configured as
an 8-channel device.
139 HP E1418A Register-Based Programming
Appendix B
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...