![HP E1418A User Manual Download Page 166](http://html.mh-extra.com/html/hp/e1418a/e1418a_user-manual_161587166.webp)
CALibrationn:CONFigure:CURRent?
CALibrationn:VALue:CURRent? <value>
Once the calibration resistance value is set as described earlier, the general
procedure for each channel current output adjustment is:
1. Make the connections shown in Figure D-5 or D-6.
2. Send the
CAL
n
:CONF:CURR?
query to the module and read the
response until a 1 is returned (typically the first query response).
3. Read the voltage output on the DMM.
4. Send the DMM value read with the
CAL
n
:VAL:CURR? <
value
>
query.
5. Enter the query response.
6. Repeat steps 3, 4, and 5 until the
CAL
n
:VAL:CURR? <
value
>
query
returns a 0.
7. Repeat steps 1 through 6 for each channel.
The first query,
CAL
n
:CONF:CURR?
, sets the channel to the calibration
mode and returns a 1 when the channel is ready for calibration. When a 1 is
returned, the channel output voltage is set to the first adjustment point.
Read the output voltage on the DMM and return the value obtained using
the
CAL
n
:VAL:CURR? <
value
>
query. This query returns an integer
indicating the calibration state of the channel. A value other than 0 or 2
from this query indicates additional values are needed.
Each channel will require multiple iterations of the
CAL
n
:VAL:CURR?
<
value
>
query. A minimum of 9 queries, to a maximum of 50 queries, will
be required at each channel. During the process, the output voltage will
range from -5 to +5 V and the last values output will be at or near 0 volts.
Storing the
Adjustments
Once all adjustments have been made using the procedures outlined
previously, you can store the new calibration constants in non-volatile
memory by sending the following command:
/* Store the new constants */
CALibration:STORe
This command will generate an error if the Cal Store Enable Jumper is not
set to the CAL position (see page 159).
Note
The HP E1418A will be busy storing the new calibration constants following
the
CAL:STORe
command. You should not perform any bus resets or module
Appendix D
Voltage/Current Output Adjustment 165
Summary of Contents for E1418A
Page 6: ...Notes 6 Contents HP E1418A 8 16 Channel D A Converter Module ...
Page 10: ...Notes 10 HP E1418A User s Manual ...
Page 12: ...12 HP E1418A User s Manual ...
Page 105: ...TRIGger 105 HP E1418A SCPI Command Reference Chapter 3 ...
Page 111: ...Notes HP E1418A Command Quick Reference 111 HP E1418A SCPI Command Reference Chapter 3 ...
Page 135: ...135 HP E1418A Register Based Programming Appendix B ...
Page 157: ...Notes 156 HP E1418A Error Messages Appendix C ...
Page 170: ...Notes Appendix D Voltage Current Output Adjustment 169 ...
Page 174: ...Figure E 1 8 Channel Disassembly 172 Configuration and Disassembly Appendix E ...
Page 175: ...Figure E 2 16 Channel Disassembly Appendix E Configuration and Disassembly 173 ...
Page 192: ...Notes 192 HP E1418A 8 16 Channel D A Converter Module Index ...