Example: Making
2-Wire Ohms
Measurements by
Scanning
Use the same setup shown in the first program example in this section,
except the current source from the multimeter must be connected to the B
tree terminal. Change the commands in lines 10 and 40 to the following:
10 OUTPUT 722;"TRIG EXT;OHM"
!Set multimeter to 2-wire ohms
40
OUTPUT 70914;"SCAN:MODE RES"
!Closes the appropriate tree
isolation switches
This command, when used with
SCAN:PORT ABUS
and
SCAN <
channel_list
>
commands, automatically closes the channels defined in the channel list, along
with the appropriate tree isolation switches. The current source is
superimposed on the channel being measured.
Example: Making
4-Wire Ohms
Measurements by
Scanning
This type of measurement requires a paired channel closure (for example,
channels 02 and 10). Use the
SCAN:MODE FRES
command to
automatically close the paired channels to the selected channels defined in
the
SCAN <
channel_list
>
command. Connect High and Low to a Bank 0
channel and High and Low to the paired channel eight numbers away. Use
only one of the channels of the channel pairs in the channel_list. Change
the commands in lines 10, 40 and 60 to the following:
10
OUTPUT 722;"TRIG EXT;OHM"
!Set multimeter to 4-wire ohms
40
OUTPUT 70914;"SCAN:MODE FRES"
!Closes the appropriate tree
isolation switches
60
OUTPUT 70914;"SCAN (@100:107,200:207)"
Figure 3-5. Mainframe to Multimeter Connection for Synchronization
38 Using the HP E1351A/53A FET Multiplexer Modules
Chapter 3
Summary of Contents for E1351A
Page 4: ...Notes 4 HP E1351A 53A FET Multiplexer User s Manual Contents ...
Page 8: ...Notes 8 HP E1351A E1353A 16 Channel FET Multiplexer Modules User s Manual ...
Page 10: ...10 HP E1351A 53A 16 Channel FET Multiplexer Module User s Manual ...
Page 28: ...Notes 28 Configuring the HP E1351A 53A FET Multiplexer Modules Chapter 2 ...
Page 78: ...Notes 78 HP E1351A 53A 16 Channel FET Multiplexer Command Reference Chapter 5 ...
Page 80: ...Notes 80 HP E1351A 53A FET Multiplexer Specifications Appendix A ...
Page 100: ...Notes 100 HP E1351A 53A FET Multiplexer Error Messages Appendix C ...