Chapter 1
Getting Started with the HP E1351A/53A
Using This Chapter
This chapter describes the HP E1351A 16-Channel FET and the
HP E1353A 16-Channel Thermocouple FET Multiplexer Modules, and
shows how to program the modules using SCPI (Standard Commands for
Programmable Instruments) commands. This chapter contains the
following sections:
•
FET Multiplexer Module Description . . . . . . . . . . . . . . . . . . Page 11
•
Multimeter Connection Points . . . . . . . . . . . . . . . . . . . . . . . . Page 12
•
Switchbox & Scanning Voltmeter Configurations . . . . . . . . Page 14
•
Digital Bus and Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . Page 15
•
Programming Language. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 16
•
Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 16
FET Multiplexer Module Description
The FET multiplexer module provides high-speed switching (multiplexing)
for up to 16 channels. The module can be used as a scanning multiplexer
for a scanning voltmeter configuration, or as a stand-alone multiplexer in a
switchbox configuration. The channels are numbered 00 to 15. Each
channel provides connections for High (H), Low (L) and Guard (G),
although only High and Low are switched. The FET multiplexer module
can switch up to 100,000 connections per second (100 K switches/sec).
The FET multiplexer module consists of a component assembly and a
terminal module. There are three different terminal modules, one for each
of the following applications: HP E1351A 16-Channel FET Multiplexer
Module, HP E1353A 16-Channel Thermocouple FET Multiplexer Module,
and HP E1352A 32-Channel Single-Ended FET Multiplexer Module. The
component assembly is the same for all three applications. For information
on the 32-Channel Single-Ended FET Multiplexer Module, see the
HP E1352A User’s Manual.
The component assembly contains the VXIbus interface, the FET switches,
the analog bus connector and the digital bus. The terminal modules provide
connection points for the individual channels, as well as monitoring points
for the tree terminals and the direct terminals. The FET multiplexer module
can be externally triggered from the VXIbus backplane or through the
digital bus handshake lines on the front of the component assembly.
Chapter 1
Getting Started with the HP E1351A/53A 11
Summary of Contents for E1351A
Page 4: ...Notes 4 HP E1351A 53A FET Multiplexer User s Manual Contents ...
Page 8: ...Notes 8 HP E1351A E1353A 16 Channel FET Multiplexer Modules User s Manual ...
Page 10: ...10 HP E1351A 53A 16 Channel FET Multiplexer Module User s Manual ...
Page 28: ...Notes 28 Configuring the HP E1351A 53A FET Multiplexer Modules Chapter 2 ...
Page 78: ...Notes 78 HP E1351A 53A 16 Channel FET Multiplexer Command Reference Chapter 5 ...
Page 80: ...Notes 80 HP E1351A 53A FET Multiplexer Specifications Appendix A ...
Page 100: ...Notes 100 HP E1351A 53A FET Multiplexer Error Messages Appendix C ...