Connecting Switchbox Channels to Direct Terminals
The direct terminals provide direct access to closed channels. There is a
direct terminal for each bank, Bank 0 and Bank 1. Closing any channel in
Bank 0 (channels 00 to 07) connects the channel to the Bank 0 direct
terminal. Closing any channel in Bank 1 (channels 08 to 15) connects the
channel to the Bank 1 direct terminal. You can isolate the two banks from
each other, and from the analog bus connector and tree terminals, with the
SCAN:PORT NONE
command. This command opens the A and B tree
isolation switches. Figure 3-1 shows how to connect channels 02 and 09 of
a single module to their respective direct terminals. Only one channel can
be closed at a time.
SCAN:PORT NONE
Opens Tree Isolation Switches.
CLOSe (@102)
Closes channel 02 on card 01.
CLOSe (@109)
Opens 02 then closes 09.
Figure 3-1. Connecting Channels 02 and 09 to their Respective
Direct Terminals
32 Using the HP E1351A/53A FET Multiplexer Modules
Chapter 3
Summary of Contents for E1351A
Page 4: ...Notes 4 HP E1351A 53A FET Multiplexer User s Manual Contents ...
Page 8: ...Notes 8 HP E1351A E1353A 16 Channel FET Multiplexer Modules User s Manual ...
Page 10: ...10 HP E1351A 53A 16 Channel FET Multiplexer Module User s Manual ...
Page 28: ...Notes 28 Configuring the HP E1351A 53A FET Multiplexer Modules Chapter 2 ...
Page 78: ...Notes 78 HP E1351A 53A 16 Channel FET Multiplexer Command Reference Chapter 5 ...
Page 80: ...Notes 80 HP E1351A 53A FET Multiplexer Specifications Appendix A ...
Page 100: ...Notes 100 HP E1351A 53A FET Multiplexer Error Messages Appendix C ...