Example: Connect
Channels 02 and 10 to
Tree Terminals for
4-Wire Ohms
Measurement
The
SCAN:MODE FRES
command separates Bank 0 and Bank 1 for 4-wire
resistance measurements or other stimulus/response testing. Current or
signals are connected to the B tree terminal and go out through a channel on
one bank. The voltage signal response comes in through a channel on the
other bank and goes to the A tree terminal and the analog bus connector.
The current source and voltage response must be on channels that are eight
numbers apart (for example, 02 and 10, 15 and 07). You only need to
specify the channel that the voltage reading is on as the FET multiplexer
automatically closes the corresponding channel eight numbers away. Figure
3-3 shows a current source on channel 10 and the voltage reading on
channel 02.
SCAN:MODE FRES
Isolates Bank 0 from Bank 1.
CLOSe (@102)
Closes channels 02 and 10. 02 is
routed to A tree terminal and 10 is
routed to B tree terminal.
Figure 3-3. Connecting Channels 02 and 10 for 4-Wire Ohms Measurement
Chapter 3
Using the HP E1351A/53A FET Multiplexer Modules 35
Summary of Contents for E1351A
Page 4: ...Notes 4 HP E1351A 53A FET Multiplexer User s Manual Contents ...
Page 8: ...Notes 8 HP E1351A E1353A 16 Channel FET Multiplexer Modules User s Manual ...
Page 10: ...10 HP E1351A 53A 16 Channel FET Multiplexer Module User s Manual ...
Page 28: ...Notes 28 Configuring the HP E1351A 53A FET Multiplexer Modules Chapter 2 ...
Page 78: ...Notes 78 HP E1351A 53A 16 Channel FET Multiplexer Command Reference Chapter 5 ...
Page 80: ...Notes 80 HP E1351A 53A FET Multiplexer Specifications Appendix A ...
Page 100: ...Notes 100 HP E1351A 53A FET Multiplexer Error Messages Appendix C ...