Troubleshooting
Troubleshooting
7-6
Troubleshooting SpO
2
Table 7-4
SpO
2
Messages
Message
Cause
Possible Solutions
SPO2 FAILURE
The unit detected a failure in
SpO
2
subsystem hardware.
This failure does not affect
other parts of the instrument.
Press
twice to
power cycle the SpO2 option.
If the error happens again,
call for service.
SPO2 SENSOR
FAIL
The sensor or adaptor cable is
broken.
Replace the sensor or adaptor
cable.
SPO2 CABLE OFF
Sensor or cable is discon-
nected.
Check SpO
2
connections
between CodeMaster XL+
and the sensor cable or adap-
tor cable.
Dashes appear on dis-
play instead of SpO
2
reading.
Can’t derive measurement
because:
•
The sensor is not on the
patient.
•
No pulse is detected.
•
The sensor is incorrectly
positioned.
•
The sensor is defective.
Check the patient for a pulse.
Reapply the sensor, and make
sure it is correctly positioned.
If it doesn’t work, replace the
sensor.
SPO2 NOISY
SIGNAL
Irregular pulse patterns.
Patient motion.
Reapply sensor.
Consider using a different
sensor site.
SPO2 LIGHT
INTERF
Too much interference from
external light.
Damage to sensor or adaptor
cable.
Reapply sensor.
Turn off lights in the room.
If other options do not work,
replace the sensor.
SPO2 LOW SIG-
NAL
Bad connection to patient, or
patient has poor perfusion.
Check the patient for poor
perfusion. Reapply disposable
and semi-disposable sensors.
Readjust reusable sensors.
Consider using a different
sensor site.
SpO
2
On/Off