Troubleshooting
Troubleshooting
7-3
If the Defibrillator does not Charge
If the defibrillator does not charge, perform the following steps.
1
Verify the proper setting of the Energy Select control.
2
If it is correct, follow these steps.
a.
Turn the Energy Select control to Off (Standby), and then back to the
desired energy setting.
b.
Press
again.
3
If the instrument remains unable to charge, turn the Energy Select control to Off
(Standby) and use a backup defibrillator.
4
Press
to print an event summary and keep any ECG strips from the
defibrillator for later evaluation.
5
Alert appropriate service personnel.
If the Defibrillator does not Deliver a Shock
If the defibrillator does not deliver a shock, perform the following steps.
DEFIB DIS-
ARMED
CodeMaster XL+ has internally dis-
charged energy and is now dis-
armed.
CHECK
RECORDER
The recorder door is open.
or
The recorder is out of paper.
Table 7-2
Defibrillator Messages
Message
Cause
Possible Solutions
Charg
Revie