Troubleshooting
Operational Checks
7-10
Keep hands clear of the paddle electrode edges. Use your thumbs
to depress the Shock buttons on the paddle handles.
3
With the paddles in their holders, grasp the paddle handles and press the Apex
paddle Shock button. Verify that the defibrillator does not discharge.
4
Release the Apex paddle Shock button, then press the Sternum paddle Shock
button. Verify that the defibrillator does not discharge.
5
Press
to place the defibrillator in sync mode.
6
Press and hold both Shock buttons. Verify that the defibrillator does not dis-
charge.
7
Press
again to remove the defibrillator from sync mode.
8
With the paddles in their holders, press and briefly hold both Shock buttons at
once.
9
The recorder will print a test report.
To check the instrument with the pads adaptor cable, perform the following steps.
1
Connect the Test Load (M1781A) to the Pads Adaptor Cable (M1750A/B).
2
Turn Energy Select to 100 joules.
3
Press Charge. Wait for the Charge Done indicators.
4
Press each Shock button independently. Verify that the defibrillator does not dis-
charge.
5
While depressing Charge, press and briefly hold both Shock buttons.
6
The recorder will print a test report.