Installation and Setup
Setup
A-9
You must depress both sides of the
key at the same time to return to the
setup/diagnostic menu.
11 To change settings in setup menu 2, select
from the setup/
diagnostic menu and repeat the above steps.
12 Turn the Energy Select control to Off (Standby) to leave setup/diagnostic mode.
You can use the factory setting for most values by selecting
from the setup/diagnostic menu.
The setting values can have critical impact on how your
defibrillator operates.
If the CodeMaster XL+ loses your configuration settings, it will
display the message
SETUP LOST
on the screen and use the factory
settings for all setup values. To clear the
SETUP LOST
message, go
to setup menu 1 or setup menu 2. The "Setup Lost" condition will
be cleared when you view the setup menus. You do not have to
change the values if you want to keep the factory settings.
Table 1-1, “Setup Menu 1 Settings,” on page A-9 and Table 1-2, “Setup Menu 2
Settings,” on page A-10 show configurable parameters on the instrument.
Table 1-1
Setup Menu 1 Settings
Setting
Choices
Description
Language
English, Dutch, Swedish, French, German,
Italian, Spanish, Finnish, Danish, Norwegian
Printed and displayed text language
Upper Alarm
Limits (UAL)
120, 140, 160, 20 to 280 in increments of 5
Upper heart rate limits LAL to 280
Lower Alarm
Limits (LAL)
40, 60, 90, 20 to 280 in increments of 5
Lower heart rate limits 20 to UAL
NOTE
ECG Size
SETUP MENU 2
RESTORE FACTORY SETTINGS
WARNING