COMMAND
:SYSTem:COMMunicate:SERial:BAUD . . . <numeric_value>
Sets or queries the baud rate.
<NUMERIC_VALUE>
RANGE
The possible BAUD rate values that can be entered for the
<numeric_value> parameter are: 1200, 2400, 4800, 9600, 14400, and
19200.
QUERY RESPONSE
Numeric data transferred as ASCII bytes in <NR1> format.
COMMENTS
•
This value is unaffected by *RST.
•
To change the Baud rate used at power-on, save the instrument
settings using the *SAV 0 command after changing the Baud rate.
RELATED
FRONT-PANEL
KEYS
Shift + Menu (BAUD > 19200 | 14400 | 9600 | 4800 | 2400 | 1200)
COMMAND
:SYSTem:ERRor?
Queries the oldest error in the Error Queue and removes that error from
the queue (first in, first out).
See Chapter 5, “Errors,” for detailed error information
QUERY RESPONSE
•
The response is in the following form:
<error_number>,“<error_description>”
•
The <error_number> is an integer in the range [-32768, 32767]. The
negative error numbers are defined by the SCPI standard; positive
error numbers are particular to this Counter. An error number value
of zero indicates that the Error Queue is empty.
•
The maximum length of the <error_description> is 255 characters.
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...