Chapter 4 Command Reference
:INITiate Subsystem
4-8
Programming Guide
4
:INITiate Subsystem
This subsystem controls the initiation of a measurement.
COMMAND
:INITiate:CONTinuous . . . <Boolean>
Sets or queries the state of continuously initiated measurements.
When CONTinuous is set to OFF, no measurements are made until
CONTinuous is set to ON or :INITiate[:IMMediate] is received. Once
CONTinuous is set to ON, a new measurement is initiated. On the
completion of each measurement, with CONTinuous ON, another
measurement immediately commences.
QUERY RESPONSE
•
Single ASCII-encoded byte, 0 or 1.
•
A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENTS
•
*RST: OFF
•
The commencement of the first measurement due to setting
:INITiate:CONTinuous to ON sets the Pending Operation Flag to
true. The Pending Operation Flag is set false by aborting a
measurement, or by the completion of the last measurement after
:INITiate:CONTinuous is set OFF.
•
With the measurements being made continuously, the :ABORt
command aborts the current measurement in progress, however, the
value of :INITiate:CONTinuous is unaffected. If CONTinuous was set
to ON prior to receiving :ABORt, it remains ON and a new
measurement begins.
•
When a single measurement is in progress (:INIT:CONT is OFF):
–
Error -213 (Init ignored) is generated and the state of INIT:CONT
is unaffected by :INIT:CONT ON.
–
Error -210 (Trigger error) is generated by INIT:CONT OFF.
NOTE
The Counter powers up with :INIT:CONT set to ON, but *RST sets
:INIT:CONT to OFF.
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...