Chapter 4 Command Reference
:MEMory Subsystem
Programming Guide
4-21
4
COMMAND
:MEMory:CLEAR[:NAME] . . . <name>
Resets the contents of the named power-correction profile to the default
configuration.
COMMENTS
•
Valid profile names: CORR1, CORR2, . . . CORR9
•
The data currently stored in non-volatile memory for the named
correction profile is discarded.
•
The default power-correction profile configuration consists of two data
points, both having loss values of zero. One of these points contains a
frequency value of 1 GHz, and the other contains a frequency value
equal to the highest frequency the Counter model can measure
(53150A: 20 GHz; 53151A: 26.5 GHz; 53152A: 46 GHz). The remaining
eight data points contain loss and frequency values of zero.
•
This is an event. There is no query form of this command.
RELATED
FRONT-PANEL
KEYS
Shift + Menu
(PWR CORR)
COMMAND
:MEMory:NSTates?
Queries the Number of available *SAV/*RCL STates in the instrument.
QUERY RESPONSE
•
Numeric data transferred as ASCII bytes in <NR1> format.
•
The value returned is 9.
•
The response value is one greater than the maximum which can be
sent as a parameter to the *SAV and *RCL commands.
COMMENTS
Query only.
RELATED
FRONT-PANEL
KEYS
None
Summary of Contents for 53150A Series
Page 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Page 2: ......
Page 10: ...Contents x Programming Guide ...
Page 11: ...1 Before You Start ...
Page 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Page 21: ...2 Command Summary A Quick Reference ...
Page 41: ...3 Programming Your Counter for Remote Operation ...
Page 101: ...4 Command Reference A Dictionary ...
Page 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Page 163: ...5 Errors ...
Page 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Page 184: ......
Page 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...