PhacoJack
R
THE Phacosystem
25
U/S HANDPIECE
CLEANING
FIGURE 17
PROCEDURE
1 DISCONNECT TUBING.
2 REMOVE U/S TIP & IRRIGATING TIP SLEEVE
FROM U/S HANDPIECE & DISCARD.
3 CONNECT SYRINGE TO ASPIRATION FITTING
OF U/S HANDPIECE AS SHOWN, FORCEFULLY
PUSH SYRINGE PLUNGER.
4 REPEAT STEP 3 WITH AIR.
5 CONNECT SYRINGE TO IRRIGATION FITTING
OF U/S HANDPIECE. FORCEFULLY PUSH
SYRINGE PLUNGER
6 REPEAT STEP 5 WITH AIR.
7 STERILIZE U/S HANDPIECE PER ACCEPTED
HOSPITAL/FACILITY PRACTICE.
WARNINGS!
* IF CLEANING IS NOT PERFORMED AFTER EACH
SURGICAL PROCEDURE, SALT & DEBRIS MAY COLLECT
ALONG THE INNER WALLS OF THE HANDPIECE. THIS
COLLECTION OF SALT & DEBRIS WILL COMPROMISE
THE SURGICAL PROCEDURE BY INCREASING THE FLUID
RESISTANCE AND LIKELY RISK IOP MAINTENANCE.
* DO NOT AUTOCLAVE ANY HANDPIECES UNTIL ALL
SALT & DEBRIS HAS BEEN REMOVED.
* NEVER USE WIRE CLEANING INSTRUMENTS OR STEEL
BRUSHES TO CLEAN ANY PART OF THE HANDPIECE.
U/S HANDPIECE
U/S TIP
TIP WRENCH
IRRIGATING TIP SLEEVE
DISCARD
WATER, DISTILLED
WATER, DISTILLED
AIR
STERILIZE
AIR