PhacoJack
R
THE Phacosystem
10
SPECIFICATIONS (CONSOLE)
FIGURE 2
ROCKER SWITCH
(POWER ON/OFF)
(POWER CORD)
REAR PANEL
29.8 cm (11.75 in)
12 Kg (26.5 Lb)
CONTROL SWITCH (SEE FIGURE 3)
VOLUME OF VACUUM
FEEDBACK
FOOTPEDAL CONNECTOR
I/A VACUUM
U/S VACUUM
VITRECTOR
VACUUM SENSE / VENT PORT
*SENSES VACUUM GENERATED
BY PERISTALTIC PUMP
*VENTS VACUUM (SEE FIGURE 3)
PERISTALTIC PUMP
IRRIGATION PINCH VALVE
VITRECTOR CONNECTOR
U/S POWER
*SETS MAXIMUM U/S
POWER
*AUTOMATIC TUNING
*40 KHZ / 60 WATTS MAX
*RANGE IS 0 - 100%
RATE
*SETS MAXIMUM PERISTALTIC PUMP RATE
*RANGE IS 10 - 40cc / MIN
*FIXED REFLUX RATE (~ 25cc / MIN)
REFLUX SWITCH
TREADLE
CAUTERY SWITCH
FOOTPEDAL (SEE FIGURE 3)
9.58 cm
(3.77in)
35.5 cm (13.96 in)
UNIVERSAL IEC POWER INPUT
120/220V
(100/240V)
50/60 Hz
PHACOJACK
CAUTERY HANDPIECE CONNECTOR
U/S HANDPIECE CONNECTOR
120V
*SETS MAXIMUM I/A &
VIT VACUUM
*RANGE IS 0 - 500mmHg
*PNEUMATIC GUILLOTINE
*RATE IS 60 - 600 CUTS / MIN
*SINGLE CUT (5)
*SETS MAXIMUM U/S VACUUM
*RANGE IS 20 - 120mmHg
CAUTERY
*SETS CAUTERY POWER
*BIPOLAR DIATHERMY, 1 MHZ
*RANGE IS 0 - 100 %
220V
GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN
CONNECTED TO A RECEPTACLE MARKED "HOSPITAL GRADE"
DO NOT USE DURING USE OF, OR PRESENCE OF, ANESTHETICS
ELECTRICA SHOCK HAZARD, HIGH VOLTAGE, DO NOT REMOVE CORD
I/A VAC
U/S VAC
MODE
VIT
CAUTERY
RATE
U/S POWER
PERCENT
PERCENT