PROCEDURE
8 INSERT IRRIGATION TUBING SET FITTING INTO U/S HANDPIECE
9 PASS DRIP CHAMBER TO CIRCULATING NURSE. DO NOT PERMIT DRIP
CHAMBER SPIKE TO TOUCH ANY NON-STERILE SURFACE WHILE HANDLING.
10 INSERT SPIKE INTO IRRIGATING SOLUTION BOTTLE. DO NOT
PERMIT DRIP CHAMBER SPIKE TO TOUCH ANY NON-STERILE
SURFACE WHILE INSERTING SPIKE INTO BOTTLE.
11 SET BOTTLE HEIGHT AS SHOWN.
12 PASS SILICONE TUBING PORTION OF IRRIGATION TUBING SET
TO CIRCULATING NURSE.
13 DEPRESS FOOTPEDAL TREADLE TO FIRST DETENT
(DETENT 1) AS SHOWN.
14 INSERT TUBING INTO PINCH VALVE.
SLIDE TUBING TO SEAT PROPERLY.
15 RELEASE FOOTPEDAL TREADLE.
I/A HANDPIECE
SYSTEM SETUP (IRRIGATION TUBING SET)
FIGURE 20
* CIRCULATING NURSE
* SCRUB NURSE
PINCH VALVE TOP VIEW
TUBING INSTALLED
SPIKE
65cm