PhacoJack
R
THE Phacosystem
34
I/A HANDPIECE
SYSTEM TEST (IOP MAINTENANCE)
FIGURE 26
PROCEDURE
1 PINCH IRRIGATION TUBING WITH CLAMP (HEMOSTAT). LOCATE CLAMP AS
CLOSE TO THE HANDPIECE AS POSSIBLE.
2 DEPRESS FOOTPEDAL TREADLE TO SECOND DETENT (DETENT 2) AS SHOWN.
3 NOTE PUMP START AND STOP MOTION. IF PUMP DOES NOT REMAIN
STOPPED OR THERE IS INTERMITTENT MOTION, STOP. DO NOT PROCEED
UNTIL PUMP REMAINS STOPPED FOR > 10 SEC.
4 RELEASE CLAMP (HEMOSTAT)
5 NOTE NO CHANGE IN TEST CHAMBER. IF COLLAPSE OCCURS, STOP.
6 REMOVE TEST CHAMBER. SYSTEM IS NOW READY FOR SURGERY.
7 RELEASE FOOTPEDAL TREADLE.
NORMAL
COLLAPSE
TROUBLESHOOTING TIPS
* MAKE SURE ALL CONNECTIONS TO THE
HANDPIECE ARE SECURE.
* AVOID SEVERE BENDS / KINKS IN ALL TUBING
* MAKE SURE THE HANDPIECE IS FREE
OF OBSTRUCTIONS.
* CHECK BOTTLE HEIGHT.
PATIENT EYE