background image

CODE: GZ0000472457

March, 2017

LAQUAtwin series (EC) Operation

1

   

  2017 HORIBA Advanced Techno Co., Ltd.

Instruction Manual (Operation)

COMPACT WATER QUALITY METER
LAQUAtwin-EC-11, LAQUAtwin-EC-22, 
LAQUAtwin-EC-33

Specifications

Items in package

Consumable parts sold separately

Part Names

Note

Press the switches 0.5 seconds or more unless
otherwise specified.

Initial Setup

Attaching/detaching the sensor

Note

Be careful not to twist the waterproof gasket.

Inserting/removing batteries

Electrode conditioning

Note

Before using the sensor for the first time or
after several days of disuse, perform electrode
conditioning. 

Perform calibration after electrode condition-
ing.

1. Place some drops of the conditioning solu-

tion into the measurement cell.

2. Wait 10 min before use.

There is no need to switch the meter ON.

3. Clean the flat sensor with running water.

Model

LAQUAtwin-EC-

11

22

33

Minimum sample volume 0.12 mL

Range and resolution (valid digits)

Conductivity

0 to 199 

S/cm:

200 to 1999 

S/cm:

2.00 to 19.99 mS/cm:

S/cm

S/cm

0.01 mS/cm

20.0 to 199.9 mS/cm:

0.1 mS/cm

TDS

0.0 to 99.9 ppm:
100 to 999 ppm:
1000 to 9990 ppm:

0.1 ppm

1 ppm

10 ppm

Calibration

Up to 2 points

Up to 3 points

Accuracy

*1

2% full scale (for each range)

2% full scale (0 to 19.99 mS/cm)

5% full scale (20.0 to 199.9 mS/cm)

Temperature display

0 to 50.0

C

Target

Electrical conductivity

Measurement 
principle

2 electrode bipolar AC
Titanium coated with Platinum black

Display

Custom (monochrome) digital LCD with 
backlight

Operating 
environment

5 to 40

C, 85% or less relative humidity 

(no condensation)

Power

CR2032 batteries (

2)

Battery life

Approx. 400 h continuous operation

*2

Outer dimen-
sions/mass

164 

 29 

 20 mm (excluding projections), 

Approx. 45 g (excluding batteries)

Main function

Temperature compensation (2%/

C fixed), 

waterproof

*3

, auto stable/auto hold, auto-

matic power OFF

*1 Accuracy is the closeness of agreement between 

the measured value and actual value of the stan-
dard solution in the measurement of the same 
standard solution as the one used for the calibra-
tion. Temperature during the calibration and mea-
surement is the same. The error of standard 
solutions and rounding error (

1 digit) are not 

included.

*2 The life period if the meter is used in the backlight

off mode. If the backlight is used, battery life will
shorten.

*3 IP67: no failure when immersed in water at a 

depth or 1 meter for 30 minutes.
Please note that the meter can not be used 
underwater

Items

Quantity

Sensor

S070

1

Meter

1

Storage case

1

Batteries

CR2032

2

Standard solution

1413 

μ

S/cm

1

12.88 mS/cm

1

Pipette

1

Conditioning solution

1

Instruction manual (Operation)

1

Instruction manual (Before use)

1

Items

Specifications

Part No.

Sensor

S070, COND

3200459672

Standard 
solution

514-22, 1413 

μ

S/cm

3999960110

514-23, 12.88 mS/cm

3999960111

Conditioning 
solution

514-20

3999960114

1

Measurement cell

2

Protection cover

3

Lithium batteries

4

MEAS switch

5

ON/OFF switch

6

CAL switch

7

Waterproof gasket

8

Strap eyelet

Attaching the sensor

1. Power OFF the meter.
2. 
Confirm that the waterproofing gasket is

clean and undamaged.

3. Slide the sensor onto

the meter so that
catch "A" on the back
of the meter fits into
hole "a" on the sen-
sor tongue as shown.

1

2

3

4

5 8

6

7

a

A

Waterproof
gasket

Detaching the sensor

1. Power OFF the meter.
2. 
Lift the sensor tongue

tip and slide the sensor
a little away from the
meter. 

3. Pull out the sensor all the way from the

meter.

Inserting the batteries

1. Power OFF the meter.
2. 
Slide both batteries into

the battery case as
shown.

Be sure to use two
CR2032 batteries, and put
them with the plus sides
(+) upwards.

Removing the batteries

1. Power OFF the meter.
2. 
Use a ball-point pen or

other tool to pry the batter-
ies out from the clips as
shown.

Battery

Battery clip

Reviews: