horiba LAQUAtwin-EC-11 Instruction Manual Download Page 3

LAQUAtwin series (EC) Operation

3

CODE: GZ0000472457

Maintenance

Appendix

Frequently asked questions

Storage

1. Clean the sensor with tap water.
2. 
Dab gently with soft tissue or cloth to

remove moisture on the sensor and meter.

Note

Especially be sure to treat the sensor gently to
prevent damaging it.

3. Close the protection cover before storing

the meter.

Temperature sensor adjustment

Temperature sensor adjustment is available on
LAQUAtwin-EC-22 and LAQUAtwin-EC-33. 
To perform accurate measurement with correc-
tion for temperature effects, follow the steps
below. Normally this is not necessary.

1. Ready a reference thermometer, and allow

the meter and reference thermometer to
reach to room temperature.

2. Set the display mode to temperature refer-

ring to " 

 Measurement display change"

(page 2).

3. Press the CAL switch.

The meter displays the setting screen for tar-
get temperature.

4. Press the MEAS switch to adjust the dis-

played temperature on the meter to match
the temperature indicated by the reference
thermometer.

Pressing the MEAS switch increases the dis-
played temperature. After the displayed tem-
perature reaches 40°C, it returns to 5°C.

5. Press the CAL switch again to apply the

displayed value to the adjustment.

The adjustment starts. The adjusted value
blinks with 

 and 

 displayed.

After the adjustment is complete, the adjusted
value stops blinking with MEAS and 

 dis-

played.

If Er4 (error display) appears, the adjustment
has failed. Retry the above steps increasing
the time spent on the step 1. 
If the adjustment repeatedly fails, the sensor
may have deteriorated. Replace the sensor
with new one.

Initializing calibration data

Initialize calibration in the following cases.

To delete the calibration data 

If the number of points for the last calibration is
uncertain.

After the sensor is replaced.

1. Press and hold the CAL and ON/OFF

switches for over 3 seconds when the
meter is switched OFF to Initialize calibra-
tion.

After a moment of all segment indication, the 
software version is displayed. 
And then, the display
changes as shown right.

2. Press the CAL switch.

All calibration data is reset.
When the initialization of cali-
bration data is complete, End
appears. 
The meter automatically switches OFF.

Initializing the settings

All setup choices are erased. The meter is reset
to the factory default values.

1. Press and hold the MEAS, CAL and ON/OFF

switches for over 3 seconds when the
meter is switched OFF to enter the initial-
ization.

After a moment of all segment indication, the 
software version is displayed. 
And then, the display
changes as shown right.

2. Press the CAL switch.

All calibration data is reset.
When the initialization of set-
tings is complete, End
appears. 
The meter automatically switches OFF.

Question

Answer

How can I 
check the 
sensor's con-
dition?

Perform 2-point calibration. If cali-
bration error occurs, the sensor has 
deteriorated. Replace the sensor.

Can I mea-
sure high or 
low tempera-
ture sam-
ples?

This meter cannot measure a sam-
ple with temperatures outside the 
meter's operating temperature range 
(5

C to 40

C).

The difference between the sample 
temperature and ambient tempera-
ture increases the measurement 
error. Perform measurement after 
the sample reaches the ambient 
temperature. 

The mea-
sured value 
does not 
change after 
changing the 
sample.

If 

 lights steadily in AH mode, the 

measured value is locked.
Press the MEAS switch to unlock the 
value. 
If the value does not change after 
unlocking, the sensor may be dam-
aged. Replace the sensor.

“Or” or “Ur” 
blinks in 
value mea-
surement.

The measured value may be out of 
the specified measurement range. 
Measure a standard solution to 
check, and if “Or” or “Ur” still blinks, 
replace the sensor.

 blinks 

during mea-
surement.

The measured temperature is not 
within the specified operating tem-
perature (5

C to 40

C).

If the ambient temperature is within 
the specified range and 

 blinks, 

replace the sensor.

The meter 
does not 
power ON.

Check that the batteries are inserted 
properly. If the battery voltage is low, 
replace them both with new ones at 
the same time.

Er4 is dis-
played during 
the calibra-
tion 

Please note that if you press the 
CAL switch in mV or temperature 
display mode, Er4 is displayed. This 
is because there is no calibration 
facility available for these modes.

Er1 is dis-
played soon 
power ON.

The internal IC in the meter may be 
defective. Perform meter initializa-
tion.
If Er1 is still displayed after the initial-
ization, the internal IC in the meter is 
defective. Replace the meter with a 
new one (the meter cannot be 
repaired).

Er2 is dis-
played right 
after power 
ON.

The internal IC in the meter is defec-
tive. Replace the meter with a new 
one (the meter cannot be repaired).

Er3 is dis-
played right 
after power 
ON.

The internal IC in the meter is defec-
tive. Replace the meter with a new 
one (the meter cannot be repaired).