background image

LAQUAtwin series (EC) Operation

4

CODE: GZ0000472457

Setup Mode

The setup mode allows the user to customize the meter to his specific needs.
To enter the setup mode, press and hold the MEAS and ON/OFF switches for over 3 seconds when the
meter is switched OFF. All the LCD segments appear and then the meter enters the setup mode.

Tip

To have the changes apply, you need to go through the entire steps from “Setup mode entry” to
“Setup completion” shown below. To leave a setting as it is, just press CAL switch in the setting.

To exit the setup mode with no change of settings, press the ON/OFF switch earlier than pressing
CAL switch in the last step but one, or the “Backlight setting” step.

Setup mode entry 

Unit setting

The display units can be changed.

TDS method setting (Only LAQUAtwin-EC-33)

The TDS method can be selected from the following options only on LAQUAtwin-EC-33.

FACt: KCl with factor adjustable from 0.4 to 1.0 (default 0.5)

442: Myron L 442 non-linear standard curve

En: European environmental standard non-linear curve

NACL: Non linear salinity curve

This step is bypassed on LAQUAtwin-EC-11 and LAQUAtwin-EC-22. 

All segment

indication (2 s)

Software ver.

indication (1 s)

ON/OFF switch
MEAS switch
CAL switch

S/cm (default)

S/m

Custom factor

(default)

442 standard

EN 27888

TDS (NaCl)

Factor setting (Only LAQUAtwin-EC-33 with the TDS method set to FACt) 

This step is bypassed on LAQUAtwin-EC-11 and LAQUAtwin-EC-22, and when the TDS method is set to
442, En, or NACL on LAQUAtwin-EC-33.

* The setting range is from F0.4 to F1.0.

In this setting, pressing the MEAS switch   
increases the displayed value. After the dis-
played value reaches F1.0, it returns to 
F0.4.

Measurement mode change

The measurement mode can be switched.

Note

The AH (auto hold) mode is applied only to
conductivity measurement.

*  Measurement display change is available in 

the AS mode. Refer to " 

 Measurement 

display change" (page 2).

Backlight setting

The backlight can be switched to ON or OFF.

Setup completion

F0.5 (default)

*

AS mode* (default) 

AH mode

Backlight ON 

(default)

Backlight OFF

END indication (1 s)

Powered OFF

For any questions regarding this product, please contact 
your local agency, or inquire from the following website. 
http://global.horiba.com/contact_e/index.htm

 

31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto, 
601-8306, Japan
http://www.horiba-adt.jp

Reviews: