Conductivity measurement
31
• Setting the temperature conversion (Default: ON, 2.00%/°C
)
The measured value of a sample that is not at 25°C can be converted to a value at the
selected temperature. To use the temperature conversion function correctly, temperature
coefficient (the rate of change per 1°C of the conductivity) must be set for each sample.
The setting of "Setting the temperature display" (page 22) is applied to the sample
temperature before the conversion.
1. Press
the
key to enter the setting mode.
2. Press the
keys to select "COND" (conductivity setting) and then press
the
key.
3. Press
the
keys to select "TEMP CF" (temperature conversion setting) and
then press the
key.
4. Press the
keys to enter the temperature coefficient and then press
the
key to confirm.
key
key
key
key
key
Enter a
number