Conductivity measurement
28
• Setting the cell constant (default: 1.000×1.0 cm
-1
)
A cell constant is unique for each conductivity cell. To measure conductivity correctly, the
cell constant of the conductivity cell must be set in the instrument.
1. Press
the
key to enter the setting mode.
2. Press the
keys to select "COND" (conductivity setting) and then press
the
key.
3. Press
the
keys to select "CELL" (cell constant setting) and then press the
key.
key
key