![horiba LAQUAact-EC110 Instruction Manual Download Page 102](http://html1.mh-extra.com/html/horiba/laquaact-ec110/laquaact-ec110_instruction-manual_2162260102.webp)
Technical note
86
■
Technical note
●
Conductivity standard values at various temperatures
Temp.
(
°C
)
Conductivity value at 25°C
84.00 (
S/cm)
1413 (
S/cm)
12.88 (mS/cm)
111.8 (mS/cm)
0
64.01
776
7.15
65.4
5
65.00
896
8.22
74.1
10
67.00
1020
9.33
83.2
15
68.00
1147
10.48
92.5
16
70.00
1173
10.72
94.4
17
71.00
1199
10.95
96.3
18
73.00
1225
11.19
98.2
19
74.00
1251
11.43
100.2
20
76.00
1278
11.67
102.1
21
78.00
1305
11.91
104.0
22
79.00
1332
12.15
105.9
23
81.00
1359
12.39
107.9
24
82.00
1386
12.64
109.8
25
84.00
1413
12.88
111.8
26
86.00
1440
13.13
113.8
27
87.00
1467
13.37
115.7
28
89.00
1494
13.62
117.7
29
90.00
1521
13.87
119.7
30
92.00
1548
14.12
121.8
31
94.00
1575
14.37
123.9
Summary of Contents for LAQUAact-EC110
Page 1: ...Handheld Conductivity Meter LAQUAact EC110 LAQUAact EC120 Instruction Manual CODE GZ0000427797...
Page 2: ......
Page 12: ...X M E M O...
Page 16: ...XIV M E M O...
Page 60: ...M E M O 44...
Page 88: ...M E M O 72...
Page 96: ...M E M O 80...
Page 104: ...M E M O 88...