![horiba LAQUAact-EC110 Instruction Manual Download Page 56](http://html1.mh-extra.com/html/horiba/laquaact-ec110/laquaact-ec110_instruction-manual_2162260056.webp)
Conductivity measurement
40
• Manual calibration
1. Clean the conductivity cell with pure water (or
deionized water) and wipe it with filter paper
or tissue paper.
Do not touch the black electrode part.
Refer to the instruction manual of the conductivity
cell for how to clean the conductivity cell.
2. Immerse the conductivity cell in the standard
solution.
Make sure the hole at the upper part of the cell is
immersed.
3. Press the
key to enter the calibration
mode.
Stability judgment starts and the COND icon
blinks. When the measured value stabilizes, the
COND icon changes from the blinking state to
the lit state.
4. Press the
keys to enter the electrical conductivity value of the
standard solution used for calibration at the measured temperature.
To change the decimal point, press the
key.
To change the unit, press the
key.
5. Press the
key or
key.
Calibration to the set standard solution value at the measured temperature is
performed. The 1st point calibration ends and the calibration history icon
corresponding to the calibration range lights, indicating that 1st point calibration is
complete.
The display returns to the measurement screen.
Summary of Contents for LAQUAact-EC110
Page 1: ...Handheld Conductivity Meter LAQUAact EC110 LAQUAact EC120 Instruction Manual CODE GZ0000427797...
Page 2: ......
Page 12: ...X M E M O...
Page 16: ...XIV M E M O...
Page 60: ...M E M O 44...
Page 88: ...M E M O 72...
Page 96: ...M E M O 80...
Page 104: ...M E M O 88...