Conductivity measurement
22
●
Setting the instrument
• Setting the temperature display
When a conductivity cell with a temperature sensor is used, or a conductivity cell without
a temperature sensor is used with a temperature electrode, the automatic temperature
measurement function can be used. During measurement, the temperature sensor
measures the temperature of the sample and displays the result on the instrument. If
automatic temperature measurement function is not used, or the temperature connector
is not connected to the instrument, the temperature set in the instrument is displayed.
1. Press
the
key to enter the setting mode.
2. Press the
keys to select "TEMP" (temperature setting) and then press the
key.
If the temperature connector (T) is connected, "ATC" (automatic temperature
compensation) appears. If the temperature connector (T) is not connected, "MTC"
(manual temperature compensation) appears.
In the case of "ATC", you can calibrate the temperature sensor (refer to "Calibrating
temperature sensor" (page 56)).