![Horiba Scientific DS-72G Instruction Manual Download Page 99](http://html1.mh-extra.com/html/horiba-scientific/ds-72g/ds-72g_instruction-manual_2162306099.webp)
79
10.3 Default Settings
10.3.1 Meter Default Settings
Item
Selection item/Setting range
Default
values
Security
Security management
function
Enable, Disable
Disable
Hold condition Hold setting mode
EXACT, NORMAL, BRIEF,
TIME, CUSTOM, OFF (Manual)
NORMAL
In selecting
"TIME"
Time setting value
2 seconds to 999 seconds
10 seconds
In selecting
"CUSTOM"
Time setting value
2 seconds to 60 seconds
10 seconds
Conductivity variation
width
1 digit to 100 digits
1 digit
Salinity variation width
0.10 PPT to 10.00 PPT
0.30 PPT
Resistivity variation width 1 digit to 100 digits
1 digit
TDS variation width
0.1 mg/L to 100.0 mg/L
100.0 mg/L
Interval
memory
Interval memory function
Enable, Disable
Disable
Time setting value
1 second to 999 seconds
30 seconds