40
The unit indication of the cell constant depends on the electrode.
Convert the unit to the one for the meter before input.
100 m
-1
= 1 cm
-1
1000 m
-1
= 10 cm
-1
10 m
-1
= 0.1 m
-1
3.2.2 COND Measurement Unit Setting
5.
Tap the left side numerical value to display the
numerical screen.
6.
Enter the numerical value written on the COND
electrode.
7.
Tap
or
to enter the digit written on the
COND electrode.
8.
Tap .
Reflect the setting.
To cancel the settings, tap
.
1.
Tap
on the right of the Unit item.
2.
Select S/m, S/cm or FIX.
The selected unit applies.
You can select S/m, S/cm or FIX (Unit is fixed at
mS/cm as the COND measurement unit.